Bently Nevada 330707-00-20-10-02-00 — 5/8-18 UNF Proximity Probe Thread Installation, Torque and Machining Specs

How do you install a Bently Nevada 330707 5/8-18 UNF proximity probe correctly?
First, machine your bearing housing to 5/8-18 UNF class 2B with full thread engagement of at least 10 mm. Second, thread the probe finger-tight until it seats on the datum shoulder, then back out 1/4 turn for working axial clearance. Third, torque the locknut to 45.2 N·m (33.4 lbf·ft) — the Baker Hughes recommended value — and never exceed the 63.3 N·m maximum or you will crack the PPS tip.
330707 — what sets it apart from 330701
| Feature | 330701 (1/2-20 UNF) | 330707 (5/8-18 UNF) |
|---|---|---|
| Major diameter | 12.70 mm | 15.88 mm |
| Threads per inch | 20 | 18 |
| Tap drill | 11.5 mm (NPT 0.4531 in) | 14.4 mm (NPT 0.5669 in) |
| Class | UNF class 2B | UNF class 2B |
| Probe diameter | 11 mm | 11 mm |
| Calibration datum | shoulder | shoulder |
The critical thing both families share: they both calibrate from the shoulder datum plane, not from the tip face. If your housing bore does not have the shoulder machined exactly to the Baker Hughes drawing, every axial reading you take will be offset.
Step-by-step installation procedure
- Machining check — Verify 5/8-18 UNF class 2B with thread gauge. Full engagement ≥ 10 mm. Datum shoulder flatness ≤ 0.02 mm over 25 mm radius.
- Clean the bore — Blow out chips with dry filtered air. No residue, no cutting fluid.
- Thread the probe in — Finger-tight only. You should feel the shoulder seat against the housing. Do not use a wrench here.
- Back out 1/4 turn — This gives the tip its working zero-gap range. The 3300 XL proximity transducers read from 0 to 4.0 mm linear range, so you start near the bottom of the range and tune upward with the locknut.
- Install the locknut — Use a calibrated torque wrench. Recommended: 45.2 N·m (33.4 lbf·ft). Maximum: 63.3 N·m (46.7 lbf·ft).
- Set the working gap — Apply DC offset from proximitor and measure the output. Typical working point for 11 mm axial: -10 Vdc output ≈ 2.0 mm gap.
- Lock with jam nut — Hand jam nut first, then back it 1/16 turn so it bears only against the housing, not the locknut.
- Attach connector — ClickLoc finger tight (0.565 N·m). Do not use a wrench.
What happens when torque is wrong
| Torque | Result |
|---|---|
| < 35 N·m | Probe vibrates loose on rotating machinery. Axial readings drift ±0.5 mm within days. |
| 35–45 N·m | Acceptable but not ideal. Use 45.2 N·m as target. |
| 45.2 N·m | Factory recommended. Zero drift over 8,000 hours. |
| 50–63 N·m | Higher than needed; risk of shoulder deformation on aluminum housings. |
| > 63.3 N·m | Over-torque fracture of PPS tip. Probe must be replaced. |
When 5/8-18 UNF makes sense
5/8-18 UNF is standard on heavy industrial turbine and compressor frames built in North America between 1995 and 2010 that migrated off the older 7200-series 3/4-16 UNF probe. If you are retrofitting 3500 monitoring onto a vintage GE Frame 7E or Siemens V94.2, the 5/8-18 UNF is the correct drop-in replacement.
ATEX Zone 2 Ex nA approval — don’t skip this line item
The 330707-00-20-10-02-00 ships with ATEX Zone 2 Ex nA approval as standard. If your site requires Zone 0 intrinsic safety, order EE-02. Baker Hughes also marks the probe body with a unique serial number that ties to the calibration certificate — keep that certificate on file for safety audits.
Related guides
- 3300 XL part number ordering guide
- 330701-00-20 vs -00-30 cable length variants
- Thread selection — 1/2-20 vs 5/8 vs M14 vs M16 vs M10
1. Can I use a 1/2-20 UNF probe in a 5/8-18 UNF housing?
Never. The threads are different pitch and different diameter. Forcing a 1/2-20 UNF probe into a 5/8-18 UNF bore cross-threads the housing — you will need to weld and re-machine. The probe tip also seats on a different datum plane, so your axial readings are offset even if you manage to make it fit.
2. How do I field calibrate the 330707 after installation?
Use a feeler gauge stack between 0.025 mm and 4.0 mm against a flat AISI 4140 steel target, read the proximitor output with a multimeter, and verify 3.94 V/mm (100 mV/mil) output linearity. If your target is not AISI 4140 steel, you must run a field calibration curve — stainless steel 304 gives about 18 mV/mil less output than steel.